Scanner Interface Software 5-21
Title Parameter Description
Data Correction
method
Data minimum
For changing the data correction method.
The recommended method is
Linearization.
For changing the Data Minimum.
Process
For viewing the result of the new setup
Reset
For resetting the image
N.R
Noise reduction
Auto shutter
Eliminate white borders around the
image.
Gain factor
This parameter moves the histogram to
the desired location on the graph and sets
the new PM gain value in the Anatom.ini
file.
Histogram
To view the histogram of the displayed
image
Save
To save the settings.
Exit
To exit the mini-viewer.
5.2.10.1 Using the Image Diagnostics Functions
1. Select the correct Sub-organ from the relevant Organ.
2. Select the desired Filter; then click Process to view the filtered image.
3. You can now right-click and drag to change contrast and intensity; or
click and drag to change zoom. The cursor changes accordingly.
Alternatively, you can use the sliders at the bottom right of the screen.
4. If the result is not satisfactory, you can click Reset to reset the image;
then repeat the procedure with another filter, until you get the best
image.
5. When the desired image quality is attained, click Save. Whenever this
part is selected for scanning under this unit, the image will be filtered at
this saved quality.